Japan Electric Meters Inspection Corporatio



Details of Accreditation Scope

Accreditation No,:0050

Name of Organization

Name: Japan Electric Meters Inspection Corporation

Accredited Laboratory

Name : Calibration Center Amagasaki
Japan Electric Meters Inspection Corporation
Address : 3-12-20 Nakoji, Amagasaki-shi, Hyogo-ken 661-0974, Japan
Telephone : +81-6-6491-5031
Facsimile : +81-6-6491-5034


Details of Accreditation Scope

Field : Electrical Calibrations

Quantity and Equipment to be Calibrated

Range

Best Measurement

Capability

DC

Voltage

Source

1V,1.018V,10V

±5ppm

1µ V~1V

±(5ppm+0.2µ V)

1V~1kV(except 1V,1.018V,10V)

±5ppm

Measuring

Equipment

1V,1.018V,10V

±5ppm

1µ V~1V

±(10ppm+0.2µ V)

1V~1kV(except 1V,1.018V,10V)

±5ppm

Resistive

Voltage

Divider

10V~1kV

100:1

±5ppm

10:1

±5ppm

1:1

±5ppm

DC

Resistance

Standard

Resistor

1mW

±10ppm

10mW

±10ppm

100mW

±5ppm

1W

±5ppm

10W

±5pm

100W

±5pm

1kW

±5pm

10kW

±5pm

100kW

±10pm

1MW

±10ppm

Measuring

Equipment

1mW ~10W

±10ppm

10W ~10kW

±5ppm

10kW ~1MW

±10ppm

1MW ~100MW

±40ppm

Direct

Current

Source

10µ A~1mA

±45ppm

1mA~ 30A

±90ppm

Measuring

Equipment

10µ A~1mA

±45ppm

1mA~30A

±90ppm

Standard

Shunt

1A~10A

±85ppm

10A~1kA

±280ppm

AC

Voltage

Source

1V~100V(40Hz~10kHz)

±100ppm

100V~ 1kV under (40Hz~10kHz)

±120ppm

Measuring

Equipment

100mV~ 100V(40Hz~10kHz)

±100ppm

100V~ 1kV(40Hz~10kHz)

±120ppm

Voltage

Transformer


3V~ 33kV(50Hz,60Hz)

Ratio error ±170ppm

Phase angle ±1.2'


33kV~ 77kV(50Hz,60Hz)

Ratio error ±240ppm

Phase angle ±1.3'


77kV~150kV(50Hz,60Hz)

Ratio error ±290ppm

Phase angle ±1.3'

AC-DC

Voltage

Comparator

1V~ 200V(40Hz~10kHz)

±120ppm

200V~ 1kV(40Hz~10kHz)

±130ppm


Details of Accreditation Scope

Field : Electrical Calibrations(Continued)

Quantity and Equipment to be Calibrated

Range

Best Measurement

Capability

Alternating

Current

Source

100mA (50Hz,60Hz)

±410ppm

100mA~10A(50Hz,60Hz)

±390pm

10A~50A(50Hz,60Hz)

±410ppm

Measuring

Equipment

100mA~1A(50Hz,60Hz)

±210ppm

1A~10A(50Hz,60Hz)

±240pm

10A~50A(50Hz,60Hz)

±410ppm

Current

Transformer

Rated primary current

100mA~150A(50Hz,60Hz)

Ratio error ±130ppm

Phase angle ±1.0'

Rated primary current

150A~3kA(50Hz,60Hz)

Ratio error ±160ppm

Phase angle ±1.4'

Rated primary current

3kA~5kA(50Hz,60Hz)

Ratio error ±210ppm

Phase angle ±1.6'

Electric

power

Watt

Converter

75V~300V

0.5A~20A(50Hz,60Hz)

Power factor 1,0.5 lag


±710ppm

Measuring

Equipment

1-Phase 2-Wire System

1-Phase 3-Wire System

75V~300V

0.5A~20A(50Hz,60Hz)

Power factor 1,0.5 lag


±1130ppm

Electric

Energy

Measuring

Equipment

3-Phase 3-Wire System

1-Phase 2-Wire System

1-Phase 3-Wire System

100V,110V

0.25A~5A(50Hz,60Hz)

Power factor 1, 0.866 lag 0.5 lag

0.5 lead


±870ppm